Multiparameter portable pH/conductivity meter contains 4 measurement modes, the meter is suitable for measuring the pH, mV, conductivity and TDS.
pH Mode:
? Multiparameter portable pH/conductivity meter is equipped with a backlit LCD display.
? 1 to 5 points calibration with automatic recognition for USA, NIST and DIN buffers.
? Automatic electrode diagnosis shows the pH slope and offset.
Conductivity/TDS Modes:
? 1 to 5 points calibration with automatic recognition for conductivity standards.
? Selectable cell constant (K=0.1/1/10), reference temperature (20/25°C), TDS factor, linear and pure water compensation modes.
Other Features:
? Automatic temperature compensation ensures accurate readings over the entire range.
? Auto-Read function senses and locks the measurement endpoint.
? Calibration due alarm prompts user to calibrate the meter regularly.
? Auto-Power Off effectively conserves battery life.
? Setup menu allows setting the number of calibration points, stability criteria, temperature unit, date and time, etc.
? Reset function automatically resumes all settings back to the factory defaults.
? Expanded memory stores or recalls up to 500 data sets.
? USB communication interface is easy to transfer data to PC.
? Multi-mode power scheme (batteries, power adapter and USB port) ensures that using the
meter smoothly.
Specifications
Model | Bante901P |
pH | |
Range | -2.000 to 20.000 pH |
Resolution | 0.001 / 0.01 / 0.1 pH, selectable |
Accuracy | ±0.002 pH |
Calibration Points | 1 to 5 points |
pH Buffer Options | USA / NIST / DIN / 2 custom buffers |
mV | |
Range | -1999.9 to 1999.9 mV |
Resolution | 0.1 / 1 mV, selectable |
Accuracy | ±0.2 mV |
Calibration Points | 1 to 5 points |
Calibration Solutions | 10 μS / 84 μS / 1413 μS / 12.88 mS / 111.8 mS |
Temperature Coefficient | 0.0 to 10.0%/°C |
Pure Water Compensation | Yes |
Reference Temperature | 20 / 25°C |
Cell Constant | K=0.1 / 1 / 10 / Custom |
TDS | |
Range | 0 to 10.00, 100.0, 1000 ppm, 10.00, 200.0 ppt |
Resolution | 0.01, 0.1, 1, automatic |
Accuracy | ±1% F.S. |
TDS Conversion Factor | 0.1 to 1.0 (default 0.5) |
Temperature | |
Range | 0 to 105°C / 32 to 221°F |
Resolution | 0.1°C/°F |
Accuracy | ±0.5°C / ±0.9°F |
Offset Calibration | 1 point |
Calibration Range | Reading ±10°C/°F |
Other Specifications | |
Temperature Compensation | 0 to 100°C / 32 to 212°F, manual or automatic |
Hold Function | Manual / Auto-read measurement endpoint |
Stability Criteria | Low / High |
Calibration Due Alarm | 1 to 31 days or off |
Auto-Power Off | 10 / 20 / 30 minutes after last key pressed |
Memory | Stores up to 500 data sets |
Output | USB communication interface |
Connector | BNC, 6-pin mini-DIN, 3.5 mm jack socket |
Display | Custom LCD (80 × 60 mm) |
Power Supply | 3 × 1.5V AA batteries or 5V DC power adapter |
Battery Life | Approximately 150 hours without backlight |
Environmental Conditions | 0 to 50°C / 32 to 122°F, relative humidity < 80% |
Dimensions | 170 × 85 × 30 mm |
Weight | 300 g |