Professional laboratory pH/ion meter contains 5 ion concentration measurement methods and water hardness measurement mode, 1 to 5 points calibration, accuracy: 0.002 pH.
pH Mode:
? Professional laboratory pH/ion meter is equipped with a 7 inch TFT display.
? 1 to 5 points calibration with automatic recognition for USA, NIST and DIN buffers.
? Solution Temperature Coefficient compensates for the pure water samples and references
the pH to 25°C.
ORP Mode:
? 1 point offset calibration allows adjusting the displayed value to a known standard.
? Relative and absolute millivolt modes ensure the reliable ORP measurements.
Ion Concentration Mode:
? 2 to 5 points calibration, including the 8 concentration points can be selected.
? Electrode management is capable of storing and recalling up to 3 electrode slopes.
? Selectable ion measurement methods (direct reading, known addition, known subtraction,
sample addition, sample subtraction) and concentration units (ppm, mg/L, mol/L, mmol/L).
Water Hardness Mode:
? 2 to 5 points calibration from low to high concentrations.
? Selectable measurement units (german degree, english degree, french degree, mmol/L
and mg/L) are used for professional water hardness measurements.
Other Features:
? Automatic temperature compensation ensures accurate readings over the entire range.
? Auto-Read function senses and locks the measurement endpoint.
? Timed Interval Readings sends the measurement data to a PC or printer.
? Limit Alarm automatically alerts when reading exceeds the specified range.
? Calibration due alarm prompts user to calibrate the meter regularly.
? Calibration log shows the date, time, calibration points, offset and slopes.
? Password protection prevents the unauthorized calibration and settings.
? Expanded memory stores or recalls up to 1000 data sets.
? Reset function automatically resumes all settings back to the factory defaults.
Specifications
Model | A130 |
pH | |
Range | -2.000 to 20.000 pH |
Resolution | 0.001 / 0.01 pH, selectable |
Accuracy | ±0.002 pH |
Calibration Points | 1 to 5 points |
pH Buffer Options | USA / NIST / DIN / 5 custom buffers |
ORP | |
Range | -2000.0 to 2000.0 mV |
Resolution | 0.1 mV |
Accuracy | ±0.2 mV |
Calibration Points | 1 point |
Measurement Modes | Relative / Absolute mV |
Ion Concentration | |
Range | 0.001 to 30000 (depending on the range of ISE) |
Resolution | 0.001, 0.01, 0.1, 1, automatic |
Accuracy | ±0.5% F.S. (monovalent), ±1% F.S. (divalent) |
Calibration Points | 2 to 5 points |
Calibration Solutions | 0.001 / 0.01 / 0.1 / 1 / 10 / 100 / 1000 / 10000 |
Measurement Units | ppm, mg/L, mol/L, mmol/L |
Measurement Methods | Direct reading, known addition, known subtraction,sample addition, sample subtraction |
Electrode Management | 1 to 3 |
Water Hardness | |
Range (Concentration) | 0.05 to 200 mmol/L |
Range (English Degree) | 0 to 1404°e |
Range (French Degree) | 0 to 2000°fH |
Range (Ca2?) | 0 to 8000 mg/L |
Range (CaCO?) | 0 to 19999 mg/L |
Range (CaO) | 0 to 11200 mg/L |
Resolution | 0.001, 0.01, 0.1, 1, automatic |
Accuracy | ±1% F.S. |
Calibration Points | 2 to 5 points |
Calibration Solutions | 0.01 / 0.1 / 1 / 10 / 100 mmol/L |
Temperature | |
Range | 0 to 105°C / 32 to 221°F |
Resolution | 0.1°C/°F |
Accuracy | ±0.5°C / ±0.9°F |
Offset Calibration | 1 point |
Calibration Range | Reading ±10°C/°F |
Other Specifications | |
Temperature Compensation | 0 to 100°C / 32 to 212°F, manual or automatic |
Stability Criteria | Standard / High-accuracy |
Measurement Modes | Continuous / Auto-read measurement endpoint |
Timed Interval Readings | 10 / 30 / 60 seconds, 10 / 30 minutes or off |
Calibration Due Alarm | 1 to 31 days or off |
High/Low Limit Alarm | Selectable |
Password Protection | 4 digits |
Memory | Stores up to 1000 data sets |
Output | USB communication interface |
Connector | BNC, 3.5 mm jack socket |
Display | 7 inch TFT LCD |
Power Supply | 12V DC power adapter |
Dimensions | 240 × 220 × 80 mm |
Weight | 1.7 kg |