  • Precision benchtop pH meter, 1 to 5 points calibration, Automatic Temperature Compensation, setup menu contains 9 options, accuracy: 0.002 pH.


    pH Mode:

    ? Precision benchtop pH meter is equipped with a large backlit LCD display.

    ? 1 to 5 points calibration with automatic recognition for USA, NIST and DIN buffers.

    ? Automatic electrode diagnosis shows the pH slope and offset.

    ? Automatic temperature compensation ensures accurate readings over the entire range.

    ? Calibration due alarm prompts user to calibrate the meter regularly.


    ORP Mode:

    ? 1 point offset calibration allows adjusting the displayed value to a known standard.

    ? Relative and absolute millivolt modes ensure the reliable ORP measurements.

    Other Features:

    ? Auto-Read function senses and locks the measurement endpoint.

    ? Setup menu allows setting the pH buffer group, number of calibration points, resolution,

      stability criteria, date and time, auto-power off, etc.

    ? Reset function automatically resumes all settings back to the factory defaults.

    ? Expanded memory stores or recalls up to 500 data sets.

    ? USB communication interface for data transfer and timed interval readings.


    Specifications

    ModelBante920
    pH
    Range-2.000 to 20.000 pH
    Resolution0.001 / 0.01 / 0.1 pH, selectable
    Accuracy±0.002 pH
    Calibration Points1 to 5 points
    pH Buffer OptionsUSA / NIST / DIN / 2 custom buffers
    ORP
    Range-1999.9 to 1999.9 mV
    Resolution0.1 / 1 mV, selectable
    Accuracy±0.2 mV
    Calibration Points1 point
    Measurement ModesRelative / Absolute mV
    Temperature
    Range0 to 105°C / 32 to 221°F
    Resolution0.1°C/°F
    Accuracy±0.5°C / ±0.9°F
    Offset Calibration1 point
    Calibration RangeReading ±10°C/°F
    Other Specifications
    Temperature Compensation0 to 100°C / 32 to 212°F, manual or automatic
    Hold FunctionManual / Auto-read measurement endpoint
    Stability CriteriaLow / High
    Calibration Due Alarm1 to 31 days or off
    MemoryStores up to 500 data sets
    OutputUSB communication interface
    ConnectorBNC, 3.5 mm jack socket
    DisplayCustom LCD (125 × 100 mm)
    Power Supply5V DC power adapter
    Environmental Conditions0 to 50°C / 32 to 122°F, relative humidity < 80%
    Dimensions210 × 188 × 60 mm
    Weight1.5 kg