Kanomax’s innovative Handheld Particle Counters (Models 3888 & 3889) set the standard of instrumentation in cleanroom / operating room monitoring and verification, filter testing, IAQ investigation, data center cleaning and many more highly specialized and highly sensitive applications. These beautifully engineered devices boast some of the most highly advanced technology in the industry. A combination of accuracy and practicality, these devices count airborne particles with high quality performance.
Features & Benefits
Model 3888 displays 3 particle sizes simultaneously (0.3, 0.5, 5.0μm) in cf or m3
Model 3889 displays 6 particle sizes simultaneously (0.3, 0.5, 1.0, 3.0, 5.0, 10.0μm) in cf or m3
Large, 4.3-inch full-color touchscreen
0.3 micron minimum sensitivity
Store up to 10,000 measurements
Compact and Light weight
Remote control software
Easy to read and interpret data due to large screen with full color and graphic display
Storage capacity for up to a week’s worth of data
Compliant with current ISO-14644
Ergonomic and comfortable for long periods of use
Grants the ability to control and monitor the instrument remotely through a PC
Easy data transfer to USB drive
Compliant with current ISO 21501-4 calibration standard
*NIST traceable calibration included
Applications
Monitor and verify cleanrooms to meet ISO 14644: 2015 cleanroom classification
Monitor surgical rooms
Filter testing
IAQ investigation
Data center cleaning
Main Unit Specifications
Product | Handheld Particle Counter | ||
Model | 3888 | 3889 | |
Particle Measurement | Sizes | 3 channels (0.3, 0.5, 5.0μm) | 6 channels (0.3, 0.5, 1.0, 3.0, 5.0, 10.0μm) |
Flow Rate | 0.1CFM (2.83L/min) Accuracy: ±5% (Compliant with JIS B9921 and ISO21501-4) | ||
Maximum Measurable Concentration | 10% for 2,000,000 particles/CF (compliant with JIS B9921 and ISO21501-4) | ||
Counting Efficiency | 50±20% (for PSL particles near the minimum measurable size) 100±10% (for PSL particles of 1.5 to 2 times as large as the minimum measurable size) (Compliant with JIS B9921 and ISO21501-4) | ||
False Count | ≦1 particle/5 minutes (Compliant with JIS B9921 and ISO21501-4) | ||
Size Resolution | ≦15% (for PSL particles near 0.3μm) (Compliant with JIS B9921 and ISO21501-4) | ||
Display | 4.3 inch color LCD, Resistive touch panel | ||
Communication | Standard | USB (Host: for printer and USB flash memory, Device: for PC) | |
Stand (option) | RS485, Ethernet, and WiFi connection points in the base/stand | ||
Recording Media | Media | Internal memory | |
Number and Format | Up to 10,000 records in CSV format | ||
Language | English, Japanese, Chinese, and Spanish | ||
Power | Internal Supply | Li-ion rechargeable battery (Replaceable) | |
External Supply | AC adapter - Input 100 to 240 V | ||
Continuous Operation Time | 5 hours (Single measurement) | ||
Operating Environment | Main Body | From 10 to 40 ℃, from 0 to 85%RH (With no condensation) | |
Probe | From 0 to 50 ℃, from 2 to 98%RH (With no condensation) | ||
Dimension | W213 X H100 X D55 mm | ||
Weight | 650g | ||
Hydrothermal Measurement (Available as an option for the 6- channel portable particle counter) | Measurement Range | (Temperature) From 0 to 50℃ (Humidity) From 2.0 to 98.0%RH | |
Display Resolution | (Temperature) 0.1℃ (Humidity) 0.1% | ||
Humidity Accuracy | ±3.0% (30 to 85%RH), ±5% (other range of humidity) | ||
Temperature Accuracy | ±0.5℃ | ||
Response Time | Approximately 60 seconds or less (90% response) |