Temperature - TC Type K (NiCr-Ni)
Measuring range | dependent on probe type used |
Accuracy | ±0.4 °C (-100 to +200 °C) ±1 °C (Remaining Range) |
Resolution | 0.01 °C |
Differential pressure (internal sensor)
Measuring range | 0 to +200 hPa |
Accuracy | ±0.3 hPa (0 to +3 hPa) ±1.5 % of mv (+3.1 to +40 hPa) ±2 hPa or ±1 % of fsv (41 to +200 hPa) |
Resolution | 0.1 hPa |
Pressure measurement (external probe)
Measuring range | 0 to 25 bar |
Accuracy | ±0.6 % of fsv (0 to +10 bar) ±0.6 % of fsv (>+10 to +25 bar) |
Resolution | 10 hPa |
General technical data
Dimensions | 219 x 68 x 50 mm |
Interface printer | Infrared interface |
Warranty | 2 years |
Measuring rate | 1 s to 24 h |
Data transfer | RS232 |
PC software | Easyheat |
Memory | 25,000 measuring values |
Weight | 600 g Approx. |