The Model 6162 is a unique anemometer designed for specialized high-temperature testing. Some of the primary applications of a mid-high temperature anemometer include: drying process control, forming process control, exhaust air measurement, and device performance testing (such as industrial ovens). Standards that this instrument apply to include ISO 13579 and NFPA 86.
Features & Benefits
Simultaneous measurement of air velocity and temperature up to 500℃ (932°F)
Two probes selections, Middle Temperature up to 200℃ (392°F) and High Temperature up to 500℃ (932°F)
Temperature compensation circuitry ensures accurate velocity measurements in rapidly fluctuating air stream temperatures
RS232C interface, analog output and remote control terminal equipped
Store up to 999 measurements
Simultaneous display of air velocity and temperature
Secondary temperature compensation circuit provides quick response time
Standard with a NIST traceable calibration certificate*
Graphic display allows easy review of historical data
6162 allows you to easily interchange smart probes
*NIST TRACEABLE CALIBRATIONS available
Applications
Drying process control
Forming process control
Exhaust air measurement
Device performance testing
ISO 13579 energy efficiency testing
NFPA 86 testing for industrial ovens
Main Unit Specifications
General Specifications | |
Analog Output | DC 0 to 1V |
Interface | RS-232C/printer |
Input | Remote start/stop function |
Datalogging | Store up to 999 measurements |
Power Supply | 6 × 1.5 V C cells / AC adapter (100 to 240 V) |
Operating Environment | |
Main Unit | 41 – 104 oF (5 – 40 oC) |
Storage Environment | 14 to 122°F (-10 to 50°C) with no condensation |
Dimensions | |
Main Unit | 8.7″ x 5.9″ x 3.3″ (220 x 150 x 85mm) |
Weight | 3.9 lbs (1.8kg) |
What’s Included | Meter, operation manual, AC adapter, analog output cable (2pc), shoulder strap, C-batteries (6 qty) |
Warranty | 1 year |
Probe Specifications
Model | 0203 | 0204 | |
Temperature Range | |||
Air Velocity Measuring Range | 32 to 212°F | 40 to 9840 fpm (0.2 to 50.0 m/s) | |
212 to 392°F | 80 to 9840 fpm (0.4 to 50.0 m/s) | ||
392 to 572°F | n/a | 138 to 9840 fpm (0.7 to 50.0 m/s) | |
572 to 752°F | n/a | 197 to 9840 fpm (1.0 to 50.0 m/s) | |
Accuracy | +/- 3% of reading | ||
Temperature Measuring Range | 32 to 392°F (0 to 200°C) | 32 to 932°F (0 to 500°C) | |
Accuracy | +/- 1% of reading | ||
General Specifications | |||
Dimensions / Weight | ? 11 x 200mm (? 0.4″ x 7.8″), 0.4 lbs (200 g) | ? 14 x 1000mm (? 0.6″ x 39.4″)1.1 lbs (500 g) | |
Probe Cables | Teflon Coating | ||
Heat-resistance | 392°F (200°C) | ||
Extension Cable | Vinyl Coating | ||
Heat-resistance | 176°F (80°C) |